- Recall classification
- Class II
Reason stated in the source
Calibration offset between the central CT image plane and the external RT lasers has been set automatically to the factory default of 50.00 cm, if the misadjustment of the offset value is not recognized during the subsequent treatment planning process or remains undetected during the position verification at the treatment system (e.g. linear accelerator), this can potentially result in an incorrect dose or dose applied to the wrong location
Source facts
- Action
- Siemens Medical Solutions USA, Inc. issued Customer Safety Advisory Letter (CSAN) SY051/20/S on August 19, 2020 via email/hand delivery by Siemens Customer Service Engineer,. The letter states reason for recall, health risk and action to take: we highly recommend performing the laser calibration and the quality assurance procedure for the laser offset after every RT Image Suite software update.Please update the calibration offset in the MM RT Image Suite Configuration dialog accordingly . Details of this procedure are documented in Chapter 13.2 (Laser configuration) of your user manual. If you have any questions or require technical support, please contact our service organization at 1-800-888-7436. The technical solution distributed via SY040/20/S. Distribution of SY040/20/S (VB40B_HF01) will be distributed to all affected customers, free of charge. Distribution of SY040/20/S began on August 26, 2020. SY040/20/S will be installed with support from your local service organization
- Initiated
- 2020-08-19
- Root cause
- Software design
- Terminated
- 2023-04-19
- Distribution
- US Nationwide distribution.
- Recall number
- Z-3005-2020
- Recall status
- Terminated
- CDRH record ID
- 182994
- FDA record created
- 2020-09-23
- Recall event number
- 86185