Skip to content

Public FDA record

Z-3006-2020: BECKMAN COULTER ACCESS Immunoassay Systems - Access Unconjugated Estriol, REF 33570.

Recall record Z-3006-2020 for Beckman Coulter Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-3006-2020
Company in FDA record
Beckman Coulter Inc.
FDA record creation date
Sep 24, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to alkaline phosphatase (ALP) associated interference causing falsely elevated uE3 (Unconjugated Estriol assay) results.

Source facts

Action
On 08/31/2020, the firm sent an "URGENT MEDICAL DEVICE RECALL" notification to customers via email and US Post informing them of the potential for an interference with bovine alkaline phosphatase (ALP) in a subset of patient samples tested with the Access Unconjugated Estriol (uE3) assay. Customers are instructed to: - Consider the patients total clinical presentation when interpreting results. - Any retrospective review of patient results is left to the discretion of the Medical Director. - Contact Beckman Coulter Technical Support if you suspect ALP associated interference and do not report the uE3 test result. - Share this information with their laboratory staff and retain this notification as part of their laboratory Quality System documentation. If they have forwarded any of the affected products to another laboratory, to please provide them with a copy of the Customer Notification letter. -Complete and return the enclosed Response Form within 10 days so that the Recalling Firm can be assured that customers have received the important communication. In the meantime, the Recalling Firm is actively working to resolve this issue by end of Q1, 2021. For any questions, please contact the Recalling Firm's Customer Support Center at: - website: http://www.beckmancoulter.com " phone: call 1-800-854-3633 in the United States and Canada. " Outside the United States and Canada contact your local representative.
Initiated
2020-08-26
Root cause
Under Investigation by firm
Terminated
2024-05-24
Distribution
Worldwide distribution - U.S. Nationwide distribution including in the states of AL, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IL, IN, KS, LA, MA, MD, MI, MN, NC, NJ, NM, NY, OH, OK, OR, PA, RI, SC, TN, TX, WA, WI and Puerto Rico. The countries of Australia, Belgium, Bosnia and Herzegovina, Brazil, Canada, China, Czech Republic, Germany, Hungary, India, Ireland, Italy, Jordan, Lebanon, Malaysia, Mexico, Nepal, Pakistan, Panama, Portugal, Russia, Saudi Arabia, Serbia, Slovakia, South Africa, South Korea, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom, and Vietnam.
Recall number
Z-3006-2020
Recall status
Terminated
CDRH record ID
183566
FDA record created
2020-09-24
Recall event number
86386

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.