Open original FDA record ↗Recall record
- FDA identifier
- Z-3010-2026
- Company in FDA record
- AVID Medical, Inc.
- FDA classification date
- Aug 21, 2026
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Labeling Deficiency: The medical kit labeling fails to identify that the statlock component contains natural rubber latex.
Source facts
- Action
- On July 9, 2026 Avid Medical ( subsidiary of Owen & Minor) issued a Urgent: Medical Device correction notification to affected consignees. Avid Medical ask consignees to take the following actions: 1. Identify, segregate, and quarantine all impacted product identified. 2. Add warning labels to affected AVID kits to ensure all users are aware of the actions to discard the impacted component prior to use. A labeling template is enclosed for printing labels which can be affixed on all affected kits and cases in inventory. 3. Customers (end-users): If you received product from a distributor, please contact the distributor or your sales representative with any questions you may have. 4. Distributors: If any of the affected kits identified above were further distributed, please ensure all end-users are appropriately notified and maintain records of effectiveness. 5. Complete the attached Recall Response Form and return by July 24, 2026.
- Initiated
- 2026-07-09
- Root cause
- Labeling Change Control
- Distribution
- US Nationwide distribution in the states of MS, TX.
- Recall number
- Z-3010-2026
- Recall status
- Open, Classified
- CDRH record ID
- 221746
- FDA classified
- 2026-08-21
- Recall event number
- 99527