Skip to content

Public FDA record

Z-3011-2026: Halyard TOTAL JOINT Model/Catalog Number: MSBM036-03

Recall record Z-3011-2026 for AVID Medical, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-3011-2026
Company in FDA record
AVID Medical, Inc.
FDA classification date
Aug 21, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Labeling Deficiency: The medical kit labeling fails to identify that the statlock component contains natural rubber latex.

Source facts

Action
On July 9, 2026 Avid Medical ( subsidiary of Owen & Minor) issued a Urgent: Medical Device correction notification to affected consignees. Avid Medical ask consignees to take the following actions: 1. Identify, segregate, and quarantine all impacted product identified. 2. Add warning labels to affected AVID kits to ensure all users are aware of the actions to discard the impacted component prior to use. A labeling template is enclosed for printing labels which can be affixed on all affected kits and cases in inventory. 3. Customers (end-users): If you received product from a distributor, please contact the distributor or your sales representative with any questions you may have. 4. Distributors: If any of the affected kits identified above were further distributed, please ensure all end-users are appropriately notified and maintain records of effectiveness. 5. Complete the attached Recall Response Form and return by July 24, 2026.
Initiated
2026-07-09
Root cause
Labeling Change Control
Distribution
US Nationwide distribution in the states of MS, TX.
Recall number
Z-3011-2026
Recall status
Open, Classified
CDRH record ID
221747
FDA classified
2026-08-21
Recall event number
99527

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗