Z-3013-2026: Midmark Digital ECG HD+12, HD+15, Model/Catalog Number: 4-000-0080 Software Version: Midmark IQecg Plugin

Recall record Z-3013-2026 for Midmark Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-3013-2026
Company in FDA record
Midmark Corporation
FDA classification date
Aug 25, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

A communication anomoly may result in incorrect electrocardiogram waveforms of a patient and an inaccurate beats-per-minute calculation.

Source facts

Action
n URGENT MEDICAL DEVICE CORRECTION notification letter dated 7/17/26 was sent to customers. Safety Instructions Until the corrected IQecg Plugin software has been installed, for electrocardiogram tracings that reflect a fast heart rate (tachycardia), Midmark recommends performing a manual clinical assessment of the patient's heart rate prior to initiating therapeutic intervention. This bedside verification should be used to confirm the accuracy of the displayed ECG heart rate before making clinical management decisions based on the device's displayed value. If this issue is encountered, follow the temporary workaround instructions provided in Appendix A. Midmark will contact your facility with additional information regarding verification of your software installation and implementation of this Field Corrective Action. Product Correction Midmark has developed a corrected version of the IQecg Plugin software that addresses the issue described in this notification and is implementing a Field Corrective Action for affected software versions. No immediate action is required to contact Midmark at this time. Midmark will contact your facility to verify the IQecg Plugin software version(s) installed on your workstations and coordinate the appropriate correction process. Depending on your facility's configuration and support requirements, the correction may be completed using Midmark-provided installation instructions or with assistance coordinated by Midmark Technical Support or an authorized service provider. Contact Information Please be assured that maintaining a high level of safety and quality is our highest priority. If you have any questions, please contact the Technical Service Team at 1.844.856.1230, option 2, or email techsupport@midmark.com.
Initiated
2026-07-20
Root cause
Software design
Distribution
Worldwide distribution - US Nationwide and the country of Canada.
Recall number
Z-3013-2026
Recall status
Open, Classified
CDRH record ID
221499
FDA classified
2026-08-25
Recall event number
99404

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗