Open original FDA record ↗Recall record
- FDA identifier
- Z-3014-2011
- Company in FDA record
- Aesculap Inc
- FDA classification date
- Aug 12, 2011
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Aesculap Implant Systems received a notification from the manufacturer, Aesculap, AG of a fractured CoCr neck adapter.
Source facts
- Action
- AESCULAP Implant Systems sent an "Important Recall Notification" letter dated to all affected customers. The informed customers of a complaint received where fractured CoCR adapter was found and to return any affected product to the firm. The reason for the fracture is currently under investigation. While the firm is trying to determine the root cause of the fracture, they will be offering the Metha non-modular short hip implants to their customers. Customers are asked to return an Inventory Sheet to AESCULAP. For additional information on this recall call (610) 984-9239 or (610) 984-9300.
- Initiated
- 2011-07-19
- Root cause
- Unknown/Undetermined by firm
- Terminated
- 2013-03-19
- Distribution
- Nationwide (USA) Distribution including the states of AL, FL, GA, MD, NM, NY, OK, and TN.
- Recall number
- Z-3014-2011
- Recall status
- Terminated
- CDRH record ID
- 102839
- FDA classified
- 2011-08-12
- Recall event number
- 59463