Skip to content

Public FDA record

Z-3014-2026: Alphenix Models INFX-8000C INFX-8000F INFX-8000V This device is a digital radiography/fluoroscopy system

Recall record Z-3014-2026 for Canon Medical System, USA, INC.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-3014-2026
Company in FDA record
Canon Medical System, USA, INC.
FDA classification date
Aug 25, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to the fixing screws of ceiling movement gear in ceiling suspended arm that are part of the X-ray interventional systems may become loose.

Source facts

Action
On 08/04/2026, an "URGENT: MEDICAL DEVICE CORRECTION" letter was sent via email informing customers that, It has been found there is a possibility that the fixing screws of ceiling movement gear in ceiling suspended arm that are part of the X-ray interventional systems may become loose. Indications that the screws have loosened are abnormal noises and sensor errors from the ceiling mounted C-arm. Customers are instructed to: Should the suspended device make an abnormal noise, or a sensor error is displayed before the corrective action is implemented on your system, please stop using the device and call your local Canon Service Team. Additionally, it is strongly requested that you share the contents of this letter with all users as well as clinical engineering or biomedical group at your facility. A representative of Canon Medical Systems USA will contact customers to schedule a time to check for the tightening and proper attachment of the fixing screws to correct this issue your system(s). For Questions - contact your service representative contact the InTouch team at intouchdispatch@us.medical.canon.
Initiated
2026-08-04
Root cause
Process control
Distribution
U.S. Nationwide distribution in the states of CA, GA, MN, NC, NV, PA, OH, and TX.
Recall number
Z-3014-2026
Recall status
Open, Classified
CDRH record ID
221087
FDA classified
2026-08-25
Recall event number
99222

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗