Z-3015-2011: Fluoroscan Mini C-arm, Models: FSIII, FSIIIC and Officemate The Officemate, FSIII and FSIIIc are compact

Recall record Z-3015-2011 for Hologic, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-3015-2011
Company in FDA record
Hologic, Inc.
FDA classification date
Aug 12, 2011
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 17, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

The Flex arm connector to C-arm can break and the C-arm that is supported by the flex arm may fall down

Source facts

Action
The firm, Hologic, Inc., sent an "Urgent: Medical Device Recall" letter dated July 25, 2011 via FedEx to its customers. The letter describes the product, problem and actions to be taken. The customers were instructed to visually check the the Flex arm/C-arm on their unit for any sign of damage or cracking, if any damage or cracks are observed, remove the unit from service until they receive the retrofit kit; and to complete and return the Confirmation of Notification form via fax to Hologic at 781-280-0662 or mail using the enclosed pre-addressed, pre-paid return envelope. After the customer confirms that they have the unit in their possession, Hologic will expedite the user installable repair kit. The kits will be available starting the first week of August. The kits will also be dispatched via FedEx. Confirmation of installation will be requested for all dispatched kits. Should you have any questions regarding this recall, require further information or assistance with the software installation call Hologic's Help Desk at 1-800-321-4659.
Initiated
2011-07-21
Root cause
Nonconforming Material/Component
Terminated
2012-06-19
Distribution
Worldwide distribution: USA (nationwide) and countries of: Argentina, Spain, and Viet Nam.
Recall number
Z-3015-2011
Recall status
Terminated
CDRH record ID
102624
FDA classified
2011-08-12
Recall event number
59398

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗