Open original FDA record ↗Recall record
- FDA identifier
- Z-3016-2020
- Company in FDA record
- Cardinal Health 200, LLC
- FDA record creation date
- Sep 25, 2020
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Only providing one nasal swab in our pre-operative kit instead of the 4 required for proper treatment.
Source facts
- Action
- On July 30th the firm send its consignee a letter notifying them of the field action. Component Part Number: X12048 (SWABSTICK, PVP-I,10%, NASAL DECOLONIZATION, PROFEND,NS) is supplied to Cardinal Health, Presource in a package of 4 nasal swabs. Per the suppliers Instructions for Use, X12048 is a kit that needs to be administered in a 4-swab treatment with two swabs used per nasal passage. Presource is only providing one nasal swab in our pre-operative kit instead of the 4 required for proper treatment. The Presource kit was set up requiring 1 each of the swab component, but confusion in the component set-up caused 1 swab to be kitted instead of 1 package of 4 swabs.
- Initiated
- 2020-07-30
- Root cause
- Employee error
- Terminated
- 2021-04-28
- Distribution
- US Nationwide distribution including the state of North Carolina.
- Recall number
- Z-3016-2020
- Recall status
- Terminated
- CDRH record ID
- 182958
- FDA record created
- 2020-09-25
- Recall event number
- 86178