Reason stated in the source
Incorrect component (1nF capacitor) was used to produce 191 printed circuit boards used in six pump lots of which 380 pumps were distributed. This issue may result in the pump displaying a "Replace Battery" alarm not related to the condition of the battery.
Source facts
- Action
- Animas Corporation sent an "Urgent: Insulin Pump Recall" letter dated July, 29 20011 to all affected customers. The letter included affected product and problem. They requested anyone who received the pump to send it back to Animas and a replacement pump will be sent at no charge. For additional information contact the Customer Support Department at (877) 937-7867.
- Initiated
- 2011-07-29
- Root cause
- Mixed-up of materials/components
- Terminated
- 2011-12-07
- Distribution
- Worldwide Distribution: Nationwide (USA) distribution including the states of CA, CO, MA, and MD; and the countries of Australia, Finland, France, Israel, Italy, Sweden and the U.K.
- Recall number
- Z-3017-2011
- Recall status
- Terminated