Open original FDA record ↗Recall record
- FDA identifier
- Z-3020-2020
- Company in FDA record
- Exactech, Inc.
- FDA record creation date
- Sep 25, 2020
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Exactech is recalling the Equinoxe Reverse Shoulder Compression Screw/Locking Cap Kit, because it may contain a incorrect length screw.
Source facts
- Action
- On 8/24/2020 the firm provided a letter to their consignees with the following instructions: In order to comply with applicable regulations and Exactech policies: " Immediately cease distribution or use of this product. " Extend this information to your accounts that may have this product in their possesrion. " Identify and quarantine any of the subject devices in your inventory (including loaer kits). " Complete and return the attached Recall Inventory Response Form to Exactech via email a recalls@exac.com.
- Initiated
- 2020-08-24
- Root cause
- Labeling Change Control
- Terminated
- 2023-05-01
- Distribution
- Domestic Distribution: KS, CA, AR, NY, OH, IL, WI, AZ, LA, KY, VA, IN, TX, HI, MN, FL, TN, GA, ID, ME, CO, WA, SC. International Distribution: Austria, Australia, Germany, France, Spain, Italy, Japan, United Kingdom, Luxembourg, Netherlands, South Africa.
- Recall number
- Z-3020-2020
- Recall status
- Terminated
- CDRH record ID
- 183624
- FDA record created
- 2020-09-25
- Recall event number
- 86396