The pumps may switch to stand-by during use requiring the pump to be reset prior to continuing treatment.
Source facts
Action
Consignees were sent a CSI "Urgent Medical Device Recall" letter dated 13 APRIL 2018. The letter listed the Affected Product, Recall Description, SIP Replacement and Return & contact information. Consignees were notified the pumps will be replaced. Consignees may continue to use their pumps however if a yellow light fault is observed and the troubleshooting per the IFU does not resolve the fault, discontinue use of the pump and notify CSI.
Initiated
2018-04-13
Root cause
Device Design
Terminated
2019-11-04
Distribution
US Nationwide Distribution in the states of AZ, CA, FL, GA, IL, MO, and TX
Recall number
Z-3021-2018
Recall status
Terminated
CDRH record ID
166793
FDA record created
2018-09-11
Recall event number
80765