An anomaly has been identified in the 4D Structure Merge option of certain versions of Varian's Eclipse Treatment Planning System. The treatment field energy may change from the current selection to a previous selection without user intervention and without invalidating the MU or calculated dose distribution.
Source facts
Action
Varian Medical Systems sent an "URGENT MEDICAL DEVICE CORRECTION/FIELD SAFETY NOTICE" letter dated July 7, 2011 to all affected customers. The letter describes the product, problem, and the actions to be taken by the customers.
The letter instructs customers to cease using the 4D Structure Merge tool option in the Eclipse until further notice. Additionally, it is recommended that customers ensure that a qualified physician reviews the accuracy and placement of all patient structures used for plan development and evaluation and to conduct standard quality assurance procedures prior to patient treatment. Customers will be contacted by a Customer Service representative when the technical correction for this issue is available.
Contact your local Varian Customer Support District or Regional Manager if you have questions regarding this notice.
Initiated
2011-07-07
Root cause
Software design
Terminated
2014-04-04
Distribution
Worldwide Distribution
Recall number
Z-3022-2011
Recall status
Terminated
CDRH record ID
102598
FDA classified
2011-08-15
Recall event number
59394