The Distal Flush Tool (DFT) accessories that were packaged in this lot number are not compatible with the TurboHawk LX-C catheter. This may result in the inability to remove excised material from the tip of the catheter.
Source facts
Action
Medtronic began notifying the Consignees on 10-Aug-2018. Consignees were asked to quarantine and return affected product.
Initiated
2018-08-10
Root cause
Process control
Terminated
2020-08-04
Distribution
Distribution to US states of KS, CA, FL, GA, AZ, MI, LA, TX, NE, KY, and China
Recall number
Z-3022-2018
Recall status
Terminated
CDRH record ID
166684
FDA record created
2018-09-11
Recall event number
80831