Complaints of this version of the power supply unit (LPS) (1.0 and 1.1) being dragged into the magnet bore (MR Scanner).
Source facts
Action
The firm sent a letter explaining the issue by email to the consignees on 08/20/2018. The firm requested that the consignee check the stock without removing the device from the magnet. It is was not in the magnet room, safely quarantine the device, and respond to the recall notice. The firm also directed any consignee who further distributed the product to notify those entities and to contact NNL. NNL will follow-up with them directly. NNL will upgrade all existing models with a safer design of the LCD Power supply unit.
Initiated
2018-08-20
Root cause
Labeling design
Terminated
2020-09-16
Distribution
worldwide
Recall number
Z-3023-2018
Recall status
Terminated
CDRH record ID
165672
FDA record created
2018-09-11
Recall event number
80529