Unexpected increase in peak tube potential (kV) will lead to an increased patient radiation dose.
Source facts
Action
Philips sent an Urgent Field Safety Notice dated August 5, 2020 to affected customers. The letter identified the affected product, problem and actions to be taken. Customers were informed that Philips plans to correct the defect at no cost to customers by implementing a field correction involving the installation of a software update. A Philips Field Service Engineers will schedule an appointment with customers to install the software update.
Initiated
2019-11-20
Root cause
Radiation Control for Health and Safety Act
Terminated
2024-04-04
Distribution
United States
Recall number
Z-3023-2020
Recall status
Terminated
CDRH record ID
183663
FDA record created
2020-09-30
Recall event number
86416