Z-3024-2018: da Vinci S/Si EndoWrist 8mm Monopolar Curved Scissors Model No. 420179-16 Product Usage: EndoWrist Instr

Recall record Z-3024-2018 for Intuitive Surgical, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-3024-2018
Company in FDA record
Intuitive Surgical, Inc.
FDA record creation date
Sep 12, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Monopolar Curved Scissor (MCS) instruments may be susceptible to developing cracks near the proximal edge of the extension tube. These cracks may fracture the plastic surrounding the instrument shaft, and may cause fragments of the extension tube to fall, including potentially into the patient.

Source facts

Action
On August 15, 2018, the firm sent an Urgent Medical Device Notification to customers via Federal Express overnight service informing them that certain lots of monopolar curved scissors may be more susceptible to developing cracks near the proximal edge of the extension tube that may cause fragments of the extension tube to fall, including potentially into patients. Customers were advised of the following 1) Always inspect the instruments prior to use for any damage; 2) Use of an instrument for a task other than the instruments designed use may result in breakage or damage; 3) During reprocessing, make certain that the MCS has been adequately rinsed and flushed, and all cleaning chemicals have been removed. Customers were instructed to read and complete actions on the firms website or to fill out and return Acknowledgement Forms to the firm. Customers with questions are encouraged to call Customer Service: - North and South America: 800-876-1310 Option 3 (4 am to 5 pm PST) - Japan: 0120-56-5635 or 03-5575-1362 (9 am to 6 pm JST) - South Korea: 02-3271-3200 (9 am to 6 pm KSTJ) - Europe, Middle East, Asia and Africa: +800 0821 2020 or +41 21 821 2020 (8 am to 6 pm CET) Field Corrective Actions: All customers who received a-16 version of S/Si MCS instrument and the -14 version of X/Xi MCS instrument will receive a customer letter, as a notification of the issue and to reiterate guidelines for proper inspection and use of the MCS instrument. Preventive Actions: The -16 version of S/Si MCS instrument and the -14 version of X/Xi MCS instruments are no longer being shipped to customers. A new design has been introduced to prevent the extension tube from being susceptible to cracking. This new material in has been shown to be robust to environmental stress cracking.
Initiated
2018-08-15
Root cause
Use error
Distribution
Worldwide Distribution - US Nationwide in the states of : AL, AK, AZ, AR, CA, CO, CT, DC, DE, FL, GA, HI, ID, IL, IN, IA, KS, KY, LA, MA, ME, MD, MI, MN, MS, MO, NB, NV, NH, NJ, NM, NY, NC, ND, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WV, WI, and WY and the countries of : Argentina, Australia, Austria, Belgium, Brazil, Canada, Chile, China, Colombia, Czech Republic, Denmark, Finland, France, Germany, Guadeloupe, Iceland, India, Ireland, Israel, Italy, Japan, Kuwait, Lebanon, Luxembourg, Martinique, Mexico, Monaco, Netherlands, Norway, Pakistan, Puerto Rico, Qatar, Romania, Russian Federation, Saudi Arabia, Singapore, Slovakia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Kingdom, and Venezuela
Recall number
Z-3024-2018
Recall status
Open, Classified
CDRH record ID
166175
FDA record created
2018-09-12
Recall event number
80696

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗