Skip to content

Public FDA record

Z-3025-2017: Discovery MR750 3.0T Product Usage: Intended use: The Discovery MR750 3.0Tsystem is a whole body magneti

Recall record Z-3025-2017 for GE Medical Systems, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-3025-2017
Company in FDA record
GE Medical Systems, LLC
FDA classification date
Aug 2, 2017
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Possible incorrect wiring of the Uninterruptable Power Supply, which could contribute to a potential shock hazard.

Source facts

Action
Consignees received a written recall notice dated 05/17/2017 notifying them of a correction to all units using an Uninterruptable Power Supply. Consignees were instructed not to unplug the unit and to continue normal use. Units with the Eaton 9130 Uninterruptable Power Supply connected to the MR operator console would receive a correction at no charge. This would involve possibly replacing the power cable. GE Representatives will be contacting consignees to arrange corrections. Any questions can be directed to GE Healthcare Service at 1-800-437-1171 or your local service representative.
Initiated
2017-05-11
Root cause
Mixed-up of materials/components
Terminated
2020-09-09
Distribution
US Nationwide in the states of : AK, CA, CO, CT, DE, FL, IA, IL, IN, KS, KY, LA, MD, MO, ND, NE, NJ, NY, OH, PA, TX, WA, and WI
Recall number
Z-3025-2017
Recall status
Terminated
CDRH record ID
156162
FDA classified
2017-08-02
Recall event number
77488

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗