Open original FDA record ↗Recall record
- FDA identifier
- Z-3025-2018
- Company in FDA record
- Intuitive Surgical, Inc.
- FDA record creation date
- Sep 12, 2018
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Monopolar Curved Scissor (MCS) instruments may be susceptible to developing cracks near the proximal edge of the extension tube. These cracks may fracture the plastic surrounding the instrument shaft, and may cause fragments of the extension tube to fall, including potentially into the patient.
Source facts
- Action
- On August 15, 2018, the firm sent an Urgent Medical Device Notification to customers via Federal Express overnight service informing them that certain lots of monopolar curved scissors may be more susceptible to developing cracks near the proximal edge of the extension tube that may cause fragments of the extension tube to fall, including potentially into patients. Customers were advised of the following 1) Always inspect the instruments prior to use for any damage; 2) Use of an instrument for a task other than the instruments designed use may result in breakage or damage; 3) During reprocessing, make certain that the MCS has been adequately rinsed and flushed, and all cleaning chemicals have been removed. Customers were instructed to read and complete actions on the firms website or to fill out and return Acknowledgement Forms to the firm. Customers with questions are encouraged to call Customer Service: - North and South America: 800-876-1310 Option 3 (4 am to 5 pm PST) - Japan: 0120-56-5635 or 03-5575-1362 (9 am to 6 pm JST) - South Korea: 02-3271-3200 (9 am to 6 pm KSTJ) - Europe, Middle East, Asia and Africa: +800 0821 2020 or +41 21 821 2020 (8 am to 6 pm CET) Field Corrective Actions: All customers who received a-16 version of S/Si MCS instrument and the -14 version of X/Xi MCS instrument will receive a customer letter, as a notification of the issue and to reiterate guidelines for proper inspection and use of the MCS instrument. Preventive Actions: The -16 version of S/Si MCS instrument and the -14 version of X/Xi MCS instruments are no longer being shipped to customers. A new design has been introduced to prevent the extension tube from being susceptible to cracking. This new material in has been shown to be robust to environmental stress cracking.
- Initiated
- 2018-08-15
- Root cause
- Use error
- Distribution
- Worldwide Distribution - US Nationwide in the states of : AL, AK, AZ, AR, CA, CO, CT, DC, DE, FL, GA, HI, ID, IL, IN, IA, KS, KY, LA, MA, ME, MD, MI, MN, MS, MO, NB, NV, NH, NJ, NM, NY, NC, ND, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WV, WI, and WY and the countries of : Argentina, Australia, Austria, Belgium, Brazil, Canada, Chile, China, Colombia, Czech Republic, Denmark, Finland, France, Germany, Guadeloupe, Iceland, India, Ireland, Israel, Italy, Japan, Kuwait, Lebanon, Luxembourg, Martinique, Mexico, Monaco, Netherlands, Norway, Pakistan, Puerto Rico, Qatar, Romania, Russian Federation, Saudi Arabia, Singapore, Slovakia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Kingdom, and Venezuela
- Recall number
- Z-3025-2018
- Recall status
- Open, Classified
- CDRH record ID
- 166178
- FDA record created
- 2018-09-12
- Recall event number
- 80696