Skip to content

Public FDA record

Z-3025-2020: Software versions syngo.CT VB20 or VB20_SP1 in the following systems: SOMATOM Force (Model #10742326) SOM

Recall record Z-3025-2020 for Siemens Medical Solutions USA, Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-3025-2020
Company in FDA record
Siemens Medical Solutions USA, Inc
FDA classification date
Sep 28, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Sporadic problems with the current software may result in scanning workflow interruptions and unexpected user notifications causing delay in diagnosis or patient rescan

Source facts

Action
Siemens Medical Solutions USA, Inc. has initiated a Customer Information Letter (CIL) to affected customers via CT004/20/S. The letter alerts customers of the technical issues and under what circumstances they may occur and health risk These issues will be resolved with software update version syngo CT VB20_SP2 (Service Pack 2) via UI CT003/20/S. The software update is being made available to all affected customers free of charge. If your system is connected to Siemens Remote Connection, this update will be performed remotely. The remote software update process will require approximately 2 hours for completion and is completely automatic; however, please make sure the system and power are stable before and during the process, please do not switch off the system during the update process. Siemens highly recommends starting the installation when the scanner will not be in use or when the necessary time for the update to be completed can be scheduled. your system is not connected to the Siemens Remote Connection, your Siemens Customer Service Engineer will schedule an appointment to perform the software update onsite. If you have any questions, please contact our service organization at 1-800-888-7436.
Initiated
2020-06-22
Root cause
Software design
Terminated
2021-03-09
Distribution
Nationwide
Recall number
Z-3025-2020
Recall status
Terminated
CDRH record ID
183480
FDA classified
2020-09-28
Recall event number
86360

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗