Skip to content

Public FDA record

Z-3026-2011: AlterG Anti-Gravity Treadmill; Alter G, 48349 Milmont Drive, Fremont, Ca 94538 Product Usage: Rehabili

Recall record Z-3026-2011 for Alter-G, Incorporated. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-3026-2011
Company in FDA record
Alter-G, Incorporated
FDA classification date
Aug 16, 2011
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 17, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Potential installation error on AlterG Anti-Gravity Treadmill that could cause greater than expected unweighting.

Source facts

Action
Alter-G, Inc. notified all affected customers by telephone, e-mail and followed by a "Medical Device Customer Advisory Notification" letter dated July 15, 2011. Customers were provided information on the installation error, the potential harm to patients, and the actions to be taken. The letter instructs customers to discontinue use of the Alter G until the recalling firm can evaluate the machine and correct any installation error. The letter states that the recalling firm is planning to visit affected customer's facility as quickly as possible and will contact customers to schedule this visit. Questions or concerns should be directed to David Tjon at 510-270-5369.
Initiated
2011-07-15
Root cause
Device Design
Terminated
2011-08-16
Distribution
Nationwide Distribution (USA ) - including the states of: OK and TX.
Recall number
Z-3026-2011
Recall status
Terminated
CDRH record ID
102672
FDA classified
2011-08-16
Recall event number
59408

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗