FDA inspection identified that due to increased positivity (false positives) on the Parvovirus B19 ELISA Test System, the firm assigned a new Correction Factor and new lot number to the remainder of the affected lot, and distributed the reworked product to the customer.
Source facts
Action
Pending. The firm has not initiated customer notification for this issue.
Initiated
2019-08-12
Root cause
Unknown/Undetermined by firm
Terminated
2022-05-20
Distribution
Worldwide distribution - US Nationwide distribution including in the states of CA, CT, IN, NJ, TX and the country of Canada.
Recall number
Z-3026-2020
Recall status
Terminated
CDRH record ID
183485
FDA record created
2020-09-29
Recall event number
86364