Z-3027-2011: Access Immunoassay Systems Total T4 Calibrators, Part Number: 33805, Subsequent Product Code: JIS Th

Recall record Z-3027-2011 for Beckman Coulter Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-3027-2011
Company in FDA record
Beckman Coulter Inc.
FDA classification date
Aug 16, 2011
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 17, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

The recall was initiated because Beckman Coulter has confirmed that the stability claims for the Access Total T4 calibrators may not be met. This issue is associated with calibrator lots 917074 and 008534.

Source facts

Action
Beckman Coulter sent an "URGENT: PRODUCT CORRECTIVE ACTION" letter dated September 13, 2010 to all affected customers. The letter describes the product, problem, and the actions to be taken by the customers. The letter instructs customers to discontinue use of the affected lot numbers and review historical quality control performance. Additionally, a response form was attached to the letter for customers to complete and return. Alternate lots will be provided for current and future replenishment orders or requests. Customers with questions regarding this notification can call Technical Support at (800) 854-3633 or contact their local Beckman Coulter representative.
Initiated
2010-08-26
Root cause
Nonconforming Material/Component
Terminated
2012-07-11
Distribution
Worldwide Distribution-USA (nationwide) and the country of Canada.
Recall number
Z-3027-2011
Recall status
Terminated
CDRH record ID
101937
FDA classified
2011-08-16
Recall event number
59239

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗