The expiration date on the device labeling exceeds its actual validated shelf life because the sterilization process used does not have adequate support for the extended shelf life.
Source facts
Action
Customers were notified via letter on about 03/02/2018. Instructions included to review inventory for affected devices, arrange for their return to SPS Sterilization, and complete and return the Buisness Response Card.
A follow up letter was sent to customer on about 03/07/2018 and included a description of the risk and hazard involved. Instructions remained the same.
Initiated
2018-03-07
Root cause
Labeling False and Misleading
Terminated
2018-11-27
Distribution
US distribution to AZ, CA, FL, OH, PA, TX, VA, WV
Recall number
Z-3027-2018
Recall status
Terminated
CDRH record ID
165303
FDA record created
2018-09-13
Recall event number
80317