Skip to content

Public FDA record

Z-3028-2020: UIBC; Catalog Number: OSR61205; UDI (01) 15099590011925; IVD; Rx Only; WARNING H316 H317 H351 H373 H411 -

Recall record Z-3028-2020 for Beckman Coulter Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-3028-2020
Company in FDA record
Beckman Coulter Inc.
FDA record creation date
Sep 30, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The manufacturer has identified that the use of Unsaturated Iron Binding Capacity (UBIC) lot 2690 will produce erroneously high patient results at the low end of the linear range.

Source facts

Action
An Urgent Medical Device Recall letter was sent on 08/18/2020 to the affected customers via mail and/or email. The customers are advised to discontinue the use of UIBC lot 2690. When available, replace it with an unaffected lot of UIBC.
Initiated
2020-08-18
Root cause
Under Investigation by firm
Terminated
2021-09-08
Distribution
Worldwide distribution - US Nationwide distribution including in the states of CA, FL and NJ. The countries of Algeria, Egypt, Philippines and Taiwan.
Recall number
Z-3028-2020
Recall status
Terminated
CDRH record ID
183255
FDA record created
2020-09-30
Recall event number
86273

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗