Open original FDA record ↗Recall record
- FDA identifier
- Z-3030-2011
- Company in FDA record
- Tosoh Bioscience Inc
- FDA classification date
- Aug 17, 2011
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
The value after reconstitution appeared to be lower than the assigned value on the label. The measured value for Lot BCAL05R was 5.3%, whereas the value on the label is 5.7%. The criteria of the value is +/- 0.3%.
Source facts
- Action
- TOSOH Bioscience sent an "Urgent Recall Notification" letter dated July, 25, 2011 to all affected customers. The letter described the problem and asks customers to immediately discontinue use of the product and return all remaining product back to the firm. Customers can contact TOSOH for replacement. For additional information or question on this recall please contact TOSOH Technical Support at (800) 248-6769.
- Initiated
- 2011-05-25
- Root cause
- Process control
- Terminated
- 2012-02-24
- Distribution
- Worldwide Distribution: Nationwide (USA) distribution including the states of FL and IL; and the country of Canada.
- Recall number
- Z-3030-2011
- Recall status
- Terminated
- CDRH record ID
- 102941
- FDA classified
- 2011-08-17
- Recall event number
- 59488