Z-3030-2026: FUJIFILM Synapse PACS software v7.4.310; radiological image processing system

Recall record Z-3030-2026 for FUJIFILM Healthcare Americas Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-3030-2026
Company in FDA record
FUJIFILM Healthcare Americas Corporation
FDA classification date
Aug 26, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Software issue may cause the ruler measurement to be inaccurate.

Source facts

Action
FUJIFILM issued an URGENT MEDICAL DEVICE CORRECTION to its consignees on 7/21/2026 via email. The notice explained the issue and how to identify it, risk to health, and requested the following: "ACTIONS TO BE TAKEN BY THE CUSTOMER/END-USER 1. Please post this notification on or near the Synapse PACs workstation. 2. Please complete and return the Field Action Verification Form attached to this communication. Responding with the information requested is essential for ensuring appropriate action is taken. 3. Please contact Technical Assistance Center (TAC) at 1-888-FUJIMED (1-888-385-4633) to open a service ticket. 4. Please report any known adverse events to FDA via MedWatch. Regarding this issue, Synapse PACS can safely be used by the end-user as long as: 5. The end-user is aware of the issue and, 6. Appropriate actions are taken when rendering a diagnosis for patients. The following workarounds can be used: 7. Perform M-mode ruler measurements on the ultrasound system and then transfer the study to Synapse PACS." For questions regarding this notice, please email hcusquality@fujifilm.com.
Initiated
2026-07-22
Root cause
Software design
Distribution
Worldwide - US Nationwide distribution in the states of NY, VA, KS, OR, OH, and the country of Australia.
Recall number
Z-3030-2026
Recall status
Open, Classified
CDRH record ID
221769
FDA classified
2026-08-26
Recall event number
99529

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗