Z-3033-2017: ThermaSure(TM), a CENORIN(TM) product, SERIES 300 MEDICAL DEVICE DRYER: (a) Model 330; (b) Model 340 For

Recall record Z-3033-2017 for Cenorin, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-3033-2017
Company in FDA record
Cenorin, LLC
FDA record creation date
Aug 14, 2017
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Dryer unit not heating. Manufacturing issue causes the terminal block to overheat, char and/or burn.

Source facts

Action
Cenorin initially contacted all domestic customers at the facility and distributor level via phone on June 7, 2017, which was followed by an Urgent Medical Device Recall letter dated July 20, 2017, to all affected customers. Cenorin sent initial notification to international distributors via email. Customers were asked to immediately examine their inventory for the affected product. If the product exhibits any symptoms of failure, immediately cease using it and contact Cenorin immediately at recall@cenorin.com or call Technical Support at 800-426-1042, ext 224.
Initiated
2017-06-07
Root cause
Process control
Terminated
2018-05-02
Distribution
Worldwide Distribution - US (Nationwide) Internationally to Japan, Hong Kong, Brunei, and Israel
Recall number
Z-3033-2017
Recall status
Terminated
CDRH record ID
157002
FDA record created
2017-08-14
Recall event number
77637

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗