Open original FDA record ↗Recall record
- FDA identifier
- Z-3034-2011
- Company in FDA record
- Interventional Spine Inc
- FDA classification date
- Aug 18, 2011
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
This recall has been initiated due to Pull Pins that may disengage from the Facet Screw during the compression step, requiring that compression be completed with a device driver rather than the compression tool. Use of this product may result in less than optimal compression leading to pseudoarthrosis, non-union or facet pain.
Source facts
- Action
- Interventional Spine, Inc. sent a recall letter dated March 15, 2011, informing consignees of the product, the problem, and the action to be taken. Consignees were asked to immediately examine their inventory and quarantine product subject to recall. In addition, if they distributed the product further they were asked to notify Jane Metcalf, the Vice President of Quality Assurance at once. Interventional Spine further informed consignees that the recall was being carried out to the individual consignee level. And their assistance was appreciated and necessary to prevent use of this product. They were then asked to complete and return the enclosed response form as soon as possible. Questions were directed to call their regional directors. For questions regarding this recall call 949-472-0006.
- Initiated
- 2011-03-15
- Root cause
- Other
- Terminated
- 2013-02-11
- Distribution
- Nationwide Distribution including CO, CT, FL, KY, LA, MA, MO, NC, NY, OK, OR, TN, TX, VA & WI
- Recall number
- Z-3034-2011
- Recall status
- Terminated
- CDRH record ID
- 99410
- FDA classified
- 2011-08-18
- Recall event number
- 58381