Open original FDA record ↗Recall record
- FDA identifier
- Z-3036-2017
- Company in FDA record
- NeuroLogica Corporation
- FDA record creation date
- Aug 25, 2017
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
When performing the Stitching function on GU60A system, unintended area is exposed to radiation
Source facts
- Action
- Neurologica will bring defect into compliance : 1.All facilities were notified via a customer notification letter. 2.Samsung will release improved software to avoid unintended exposure. When setting the stitching area, if there is an over exposed area, the height of the swivel arm will be lowered over the exposed area and the first region (first image) will be collimated by the bottom blade. 3.A software patch was developed: SW3.01 was released on 05/08/2017 and SW1.00 was released on 07/28/17. 4.The software patch will be installed on the affected devices. 5.Samsung has provided the statement and assurance that all corrections will be performed free of charge. 6.All units will be provided with updated software by November 11, 2017. For further questions, please call (978) 564-8500.
- Initiated
- 2017-05-24
- Root cause
- Device Design
- Terminated
- 2018-07-09
- Distribution
- Nationwide Distribution
- Recall number
- Z-3036-2017
- Recall status
- Terminated
- CDRH record ID
- 155900
- FDA record created
- 2017-08-25
- Recall event number
- 77384