Open original FDA record ↗Recall record
- FDA identifier
- Z-3044-2024
- Company in FDA record
- MEDLINE INDUSTRIES, LP - Northfield
- FDA classification date
- Sep 6, 2024
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Medical convenience kits contain plastic syringes affected by the FDA Safety Alert issued on March 19, 2024. Leaks, breakage, and/ other quality issues have been identified that may pose a risk to patient health.
Source facts
- Action
- Medline Industries LP issued a recall notice to its consignees on 04/08/2024 via letter delivered via USPS and email. The notice explained the issue, provided a link to for information pertaining to the risk involved https://www.fda.gov/medical-devices/safety-communications/update-evaluating-plastic-syringes-made-china-potential-device-failures-fda-safety-communication) and request these syringes within the convenience be discarded, except for circumstances where no alternative syringes are readily available. If these syringes must be used, closely monitor for leaks, breakage, and any other problems, and report any quality issues to Medline and the FDA. Distributors were directed to notify their customers. The firm issued a second recall notice (an expansion) on 05/02/2024 using the same methods.
- Initiated
- 2024-04-08
- Root cause
- Under Investigation by firm
- Distribution
- Worldwide distribution.
- Recall number
- Z-3044-2024
- Recall status
- Open, Classified
- CDRH record ID
- 208801
- FDA classified
- 2024-09-06
- Recall event number
- 94583