- FDA identifier
- Z-3046-2011
- Company in FDA record
- Steris Corporation
- FDA classification date
- Aug 22, 2011
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
During an inspection the firm was notified that language contained in labeling and promotional literature directs Customers that the products can be used to replace biological indicators and/or release loads containing implants.
Source facts
- Action
- STERIS sent a Customer Notification letter dated March 16, 2010, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to review and follow the Challenge Packs' revised instructions for use enclosed as Attachment A (270-4) and Attachment B (275F 3). For any questions regarding this recall customers were instructed to contact their STERIS Account Manager or call 440-392-7421.
- Initiated
- 2011-03-16
- Root cause
- Labeling False and Misleading
- Terminated
- 2013-03-19
- Distribution
- Worldwide Distribution - USA (nationwide) and the countries of Lebanon and Qatar.
- Recall number
- Z-3046-2011
- Recall status
- Terminated
- CDRH record ID
- 103396
- FDA classified
- 2011-08-22
- Recall event number
- 59713