- Recall classification
- Class II
Reason stated in the source
Potential inaccurate calibration by service depot repair, which could contribute to patients being burnt by the device. This only impacts units serviced between the dates of 09/08/2016 and 05/23/2017.
Source facts
- Action
- Consignees were telephoned on June 1, 2017, as initial recall notification. At that time, all consignees agreed to discontinue use of the affected product until it was returned or repaired. An Urgent Medical Device Recall letter dated June 14, 2017, requested that consignees that have not been contacted regarding this recall and have a device repaired between the dates of 09/08/2016 and 05/23/2017 contact Solta Medical immediately at 877-782-2286. Any questions regarding the recall can be directed to Solta Medical Product Support Team at 877-782-2286 or 510-259-5299, option 2.
- Initiated
- 2017-06-01
- Root cause
- Equipment maintenance
- Terminated
- 2020-11-30
- Distribution
- US Distribution to the states of : AL, CA, CO, FL, NY, PA, & TX.
- Recall number
- Z-3047-2017
- Recall status
- Terminated
- CDRH record ID
- 157186
- FDA record created
- 2017-08-21
- Recall event number
- 77655