Skip to content

Public FDA record

Z-3047-2026: STA Staclot Protein S, Model/Catalog Number: 00746-Reagent Lot number 273074; qualitative and quantitati

Recall record Z-3047-2026 for Diagnostica Stago, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-3047-2026
Company in FDA record
Diagnostica Stago, Inc.
FDA classification date
Aug 28, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

A positive drift in QC values within the reagent's stated 4-hour onboard stability period was identified

Source facts

Action
Diagnostica Stago issued an URGENT: MEDICAL DEVICE RECALL (REMOVAL) notice to its consignees on 7/29/2026 via email. The notice described the issue, risk assessment, and requested the following: "Actions Required To prevent potential workflow disruptions associated with the increased recalibration frequency, we request that you: " Immediately discontinue use of STA-Staclot Protein S, Lot 273074. " Dispose of any remaining inventory of the affected lot in accordance with your facility's procedures and local regulations. " Complete, sign, and return the enclosed Acknowledgement Form by email to confirm receipt and understanding of this notification. Stago will provide a credit for the destroyed kits you indicate on the Acknowledgement Form." For questions regarding this notice or require assistance, please contact Stago's 24/7 Hotline at 1-800-725-0607.
Initiated
2026-07-29
Root cause
Under Investigation by firm
Distribution
US Nationwide distribution in the states and territory of NJ, NY, NJ, VA, PA, FL, AL, GA, TN, OH, IL, MO, NE, TX, CO, UT, CA, PR.
Recall number
Z-3047-2026
Recall status
Open, Classified
CDRH record ID
221774
FDA classified
2026-08-28
Recall event number
99531

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.