When using the HD120 Multi Leaf Collimator for beam shaping, and under specific treatment conditions, dose may be delivered outside of the intended treatment field that exceeds the expected transmission dose.
Source facts
Action
Varian Medical Systems sent an "URGENT MEDICAL DEVICE CORRECTION" letter dated June 6, 2011 to all affected customers. The letter describes the product, problem, and the actions to be taken by the customers.
***12-20-12, AMENDED INFO ***: An additional letter was sent to consignees on 11/5/12 that Varian identified as also being affected by the recall.***
The letter describes various jaw position recommendations to be followed when planning and treating dynamic fields on the HD120 MLC.
Contact the Varian Oncology Help Desk at 1-888-827-4265 for questions regarding this notice.
Initiated
2011-06-06
Root cause
Device Design
Terminated
2011-12-23
Distribution
Worldwide Distribution.
Recall number
Z-3052-2011
Recall status
Terminated
CDRH record ID
101854
FDA classified
2011-08-22
Recall event number
59226