Z-3062-2024: Guidewire 70cm x 1mm (0.038 ) REF 5583705 These accessories are intended for use with Bard 14.5 Fr. Long

Recall record Z-3062-2024 for Bard Peripheral Vascular Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-3062-2024
Company in FDA record
Bard Peripheral Vascular Inc
FDA classification date
Sep 6, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to guidewires being manufactured incorrectly with the guidewires being mis-assembled within the hoop such that the proximal (stiff) end of the guidewire is located at the distal end of the hoop. Injury could occur if the user attempts to insert the proximal end into the patient.

Source facts

Action
On 08/01/2024, the firm sent an "URGENT: Medical Device Product Removal" letter via US mail and email to customers informing them that the firm has received multiple reports that guidewire may have been mis-assembled within the hoop such that the proximal (stiff) end of the guidewire is located at the distal end of the hoop. Customers are instructed to: 1. Discontinue use of product listed in the letter, and immediately dispose of all affected product remaining in their possession in accordance with their local facility s process. 2. Share this notification with all users within their facility network of the product to ensure they are also aware of this field action. If the affected products have been distributed further, ensure that these entities are contacted and informed of the recall. 3. Complete and return the attached Customer Response Form and return it to the BD contact noted on the form whether or not they have any of the product so that BD may acknowledge your receipt of this notification and subsequently process any required credits. 4. Indicate on the response form the quantity of product identified at their facility and confirm that this product inventory was disposed of. 5. If products were purchased from a distributor, contact the distributor for further instructions and credit resolution. For questions/assistance contact: BD North American Regional Complaint Center for Product Complaints/Technical Questions at: Phone: 1-844-8BD-LIFE (1-844-823-5433) Say Complaints when prompted Mon Fri 8:00am and 5:00pm CT, or Email: productcomplaints@bd.com. For recall related questions email BDRC14@BD.COM
Initiated
2024-07-10
Root cause
Under Investigation by firm
Distribution
Worldwide - U.S. Nationwide distribution in the states of OH, TX, and WV. The countries of Belgium, Canada, France, Italy, and United Kingdom.
Recall number
Z-3062-2024
Recall status
Open, Classified
CDRH record ID
209176
FDA classified
2024-09-06
Recall event number
95039

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗