Skip to content

Public FDA record

Z-3064-2026: HEALON EndoCoat PRO, REF: 10300012 (US/CAN) HEALON DUET PRO II, REF: 10300016 (US/CAN) HEALON EndoCoat

Recall record Z-3064-2026 for Johnson & Johnson Surgical Vision, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-3064-2026
Company in FDA record
Johnson & Johnson Surgical Vision, Inc.
FDA classification date
Aug 31, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Surgical instructions sent for viscoelastic products used during ophthalmic viscosurgical cataract procedures including in part: If resistance is encountered during insertion, withdraw cannula and reassess the insertion conditions (i.e. insertion angle, incision size, etc.) before proceeding because forcing the cannula may cause corneal abrasion, iris trauma or descemet's membrane detachment.

Source facts

Action
On 7/31/2026, correction notices were mailed to customers and distributors informing them of the following: Device Insertion Technique Clarification: If resistance is encountered during insertion, withdraw the cannula and reassess the insertion conditions (i.e. insertion angle, incision size, etc.) before proceeding. Do not force the cannula through the incision, as this may increase the risk of unintended tissue injury, including, but not limited to corneal abrasion, iris trauma or Descemet s Membrane Detachment, which may require medical or surgical intervention to preclude serious injury. This investigation is currently ongoing, and we will notify users if any additional information comes to light. 1) Forward this notice to all surgeons using this device in your facility. 2) If any of the subject product has been transported to another facility/surgeon, contact that facility/surgeon and provide/forward this notice to them. 3) Keep a copy of this notice with the subject product. 4) Complete, sign, and return the Business Reply Form via email to RegCompliOne@its.jnj.com If you have questions or concerns with regard to this notification, please contact the firm's Global Post Market Safety Team at 1-877-266-4543, Option 2. Additional notices with additional affected devices to be sent in the future.
Initiated
2026-07-31
Root cause
Device Design
Distribution
Worldwide - US Nationwide distribution including in the states of PA, AL, AZ, AR, CA, FL, IL, IN, IA, KY, LA, ME, MI, MN, MS, NJ, NY, NC, OH, OR, TN, TX, WA, WV, HI, CO, MD, VA, WI, ID, NH, SD, OK, MA, NM, DE, GA, SC, CT, NV, MO, RI, KS, ND and the countries of CA, BE, CH, DE, DK, ES, FR GB, HU, IS, IT, NL, NO, PT, SE.
Recall number
Z-3064-2026
Recall status
Open, Classified
CDRH record ID
221673
FDA classified
2026-08-31
Recall event number
99496

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗