Z-3065-2024: Dimension Vista Total Bilirubin (TBIL) Flex Reagent Cartridge; Siemens Material Number (SMN): 10445146;

Recall record Z-3065-2024 for Siemens Healthcare Diagnostics, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-3065-2024
Company in FDA record
Siemens Healthcare Diagnostics, Inc.
FDA classification date
Sep 6, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

During manufacturing, the last three tests from the 160 test flex of lot 23206BA (wells 8 or 10) were processed on the Dimension Vista platform. There was an issue with the filling line which led to some wells to underfill in lot 23206BA. Calibrator, quality control (QC) and patient results could have been impacted.

Source facts

Action
On July 30, 2024 URGENT MEDICAL DEVICE CORRECTION letters were sent to consignees. Impact to Results: 1. Erroneously depressed or elevated total bilirubin patient results may have occurred, even if QC was within range. 2. A negative bias of >-98% (erroneous result <0.1 mg/dL [2 mol/L] at a true concentration of 4.0 mg/dL [69 mol/L]) and a positive bias of up to 3,950% (erroneous result 16.2 mg/dL [277 mol/L] at a true concentration of 0.4 mg/dL [7 mol/L]) was observed. Results of this test should always be interpreted in conjunction with the patient s medical history, clinical presentation and other findings. Customer Actions: 1. Please review this letter with your Medical Director to determine the appropriate course of action, including for any previously generated results, if applicable. 2. Complete and return the Field Correction Effectiveness Check form attached within 30 days and indicate product replacement needs if retesting of patients is performed. 3. Please retain this letter with your laboratory records and forward this letter to those who may have received this product. Resolution: The cause of the filling issue has been determined and resolved by Siemens Healthineers. Mitigation that will prevent this issue from occurring in the future has been put in place. We apologize for the inconvenience this situation may cause. If you have any questions, please contact your Siemens Healthineers Customer Care Center or your local Siemens Healthineers technical support representative.
Initiated
2024-07-30
Root cause
Software Manufacturing/Software Deployment
Distribution
Domestic: Nationwide Distribution. International: Austria, Belgium, Canada, Czech Republic, Denmark, France, Germany, Italy, Qatar, Slovenia, Spain, Switzerland.
Recall number
Z-3065-2024
Recall status
Open, Classified
CDRH record ID
209481
FDA classified
2024-09-06
Recall event number
95116

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗