Error in reprocessing instructions in Revisions 13 and 14 of IFU Z845, used with EG-3670URK and EG-3870UTK ultrasound gastroscopes. In the disinfection step, the IFU has a warning indicating the cleaning detergent solution should remain in contact with the all internal channels and external endoscope surfaces. The disinfecting solution, not the cleaning detergent, should be used for this step.
Source facts
Action
Recall Notifications were sent to 104 consignees via USPS Certified Mail on 5/23/2017. Included with the Field Safety letter was the customer response form, and the addendum to IFU Z845. These were also posted on Pentax of America web page until the recall is closed.
Initiated
2017-05-23
Root cause
Error in labeling
Terminated
2020-05-21
Distribution
Nationwide
Recall number
Z-3067-2017
Recall status
Terminated
CDRH record ID
155880
FDA record created
2017-08-23
Recall event number
77380