Skip to content

Public FDA record

Z-3068-2017: Ultrasound Video Gastroscope, Model EG-3870UTK. The product is intended to be used in combination with t

Recall record Z-3068-2017 for Pentax of America Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-3068-2017
Company in FDA record
Pentax of America Inc
FDA record creation date
Aug 23, 2017
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Error in reprocessing instructions in Revisions 13 and 14 of IFU Z845, used with EG-3670URK and EG-3870UTK ultrasound gastroscopes. In the disinfection step, the IFU has a warning indicating the cleaning detergent solution should remain in contact with the all internal channels and external endoscope surfaces. The disinfecting solution, not the cleaning detergent, should be used for this step.

Source facts

Action
Recall Notifications were sent to 104 consignees via USPS Certified Mail on 5/23/2017. Included with the Field Safety letter was the customer response form, and the addendum to IFU Z845. These were also posted on Pentax of America web page until the recall is closed.
Initiated
2017-05-23
Root cause
Error in labeling
Terminated
2020-05-21
Distribution
Nationwide
Recall number
Z-3068-2017
Recall status
Terminated
CDRH record ID
155881
FDA record created
2017-08-23
Recall event number
77380

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗