Skip to content

Public FDA record

Z-3068-2024: Brand Name: Transcend Product Name: Transcend Singles Universal Body (UB) Refill 10pk Model/Catalog Numbe

Recall record Z-3068-2024 for Ultradent Products, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-3068-2024
Company in FDA record
Ultradent Products, Inc.
FDA classification date
Sep 6, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to a manufacturing issue, red particulates present in the expressed composite.

Source facts

Action
On 08/6/2024, the firm sent an "URGENT: MEDICAL DEVICE CORRECTION" Letter via 3-day delivery to customer informing them that firm has identified a potential presence of red or blue polyvinyl particulate in the composite. Customer are instructed to: -Discard all remaining Transcend Singles or Vit-lescence Singles from lots listed in the Recall Letter. -Fill out Customer Response Form and return to Ultradent indicating disposal of product. -For any concerns or questions related to your replacement product or product use, contact Ultradent Customer Service at 1-888-230-1420. -Should you experience any adverse reactions or quality problems with the use of this product, these can also be reported to the FDA s MedWatch Adverse Event Reporting program with online, by regular mail or by fax. o Complete and submit the report online o Regular mail or fax: download form or call 1-800-332-1088 to request a reporting form, then complete and return to the address on the pre-addressed form, or submit by fax to 1-800-FDA-0178
Initiated
2024-08-02
Root cause
Material/Component Contamination
Distribution
US: AZ, CA,CO,FL,GA,HI, IA, ID,IL,IN,KS,KY,LA, MA, MD, MI, MN, MO, MS, NC, ND, NE, NJ, NV, NY, OH, SC, TN, TX, UT, VA, VT, WI, WV Foreign: NL, DE, CH, CZ, FR, ES
Recall number
Z-3068-2024
Recall status
Open, Classified
CDRH record ID
209675
FDA classified
2024-09-06
Recall event number
95175

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗