Open original FDA record ↗Recall record
- FDA identifier
- Z-3070-2011
- Company in FDA record
- Arjo, Inc. dba ArjoHuntleigh
- FDA classification date
- Aug 24, 2011
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
The Entroy Pool Lift boom may rapidly lower un-commanded to the bottom of its stroke, with a potential for injury to the resident or caregiver.
Source facts
- Action
- ArjoHuntleigh sent an Urgent Field Correction Notice I.D. 110701AHE to the affected end user via UPS 2nd Day on August 4, 2011. The notice informed the account of the potential for lift boom to rapidly lower un-commanded to the bottom of its stroke. The account was requested to immediately cease use of the affected lift, and disconnect the power to the lift to prevent further use until corrected. The account was requested to complete and sign the Customer Response Form, and return it to ArjoHuntleigh to schedule a service call to correct the lift. Questions were directed to the ArjoHuntleigh at 1-800-323-1245, ext 54485.
- Initiated
- 2011-08-04
- Root cause
- Device Design
- Terminated
- 2011-11-28
- Distribution
- Worldwide Distribution - USA including CT
- Recall number
- Z-3070-2011
- Recall status
- Terminated
- CDRH record ID
- 103152
- FDA classified
- 2011-08-24
- Recall event number
- 59588