Open original FDA record ↗Recall record
- FDA identifier
- Z-3076-2011
- Company in FDA record
- Hamamatsu Corporation
- FDA classification date
- Sep 13, 2011
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
It was discovered about a potential issue of X-ray leakage with our Model # LI1 0951-35 Micro Focus X-Ray generator.
Source facts
- Action
- Hamamatsu Corporation sent a letter dated July 19, 2011, to the manufacturer and their subcontractor informing them of the affected product and what they propose to do. They will repair the units both in the field and those returned to our plant by installing additional lead shielding in the units. The lead shielding will be fixed to the unit by soldering the addition to the existing lead shield (see attachment 2). We expect to complete the repair of all units in the field by October 1, 2011. All new units to be shipped from our plant will be repaired before shipment. For further questions, please call (908) 231-0960 X 2300
- Initiated
- 2011-07-07
- Root cause
- Radiation Control for Health and Safety Act
- Terminated
- 2018-01-29
- Distribution
- Worldwide Distribution -- USA ( nationwide ) the countries of Aichi, India and Japan.
- Recall number
- Z-3076-2011
- Recall status
- Terminated
- CDRH record ID
- 103604
- FDA classified
- 2011-09-13
- Recall event number
- 59760