Skip to content

Public FDA record

Z-3076-2011: Hamamatsu model L10951-35 Micro Focus x-ray generator. The x-ray generator is incorporated into a cabine

Recall record Z-3076-2011 for Hamamatsu Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-3076-2011
Company in FDA record
Hamamatsu Corporation
FDA classification date
Sep 13, 2011
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

It was discovered about a potential issue of X-ray leakage with our Model # LI1 0951-35 Micro Focus X-Ray generator.

Source facts

Action
Hamamatsu Corporation sent a letter dated July 19, 2011, to the manufacturer and their subcontractor informing them of the affected product and what they propose to do. They will repair the units both in the field and those returned to our plant by installing additional lead shielding in the units. The lead shielding will be fixed to the unit by soldering the addition to the existing lead shield (see attachment 2). We expect to complete the repair of all units in the field by October 1, 2011. All new units to be shipped from our plant will be repaired before shipment. For further questions, please call (908) 231-0960 X 2300
Initiated
2011-07-07
Root cause
Radiation Control for Health and Safety Act
Terminated
2018-01-29
Distribution
Worldwide Distribution -- USA ( nationwide ) the countries of Aichi, India and Japan.
Recall number
Z-3076-2011
Recall status
Terminated
CDRH record ID
103604
FDA classified
2011-09-13
Recall event number
59760

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗