Open original FDA record ↗Recall record
- FDA identifier
- Z-3078-2011
- Company in FDA record
- Roche Diagnostics Operations, Inc.
- FDA classification date
- Aug 29, 2011
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Roche has detected a problem with the MagNA Pure LC 1.0 (software version 3.0.11) and the MagNA Pure LC 2.0 (software versions 1.1.23 and 1.1.24) instruments. Under certain circumstances, only a fraction of each sample is processed which may lead to false negative results, and an impact to patient health if used in a homebrew diagnostic application.
Source facts
- Action
- Roche Diagnostics Operations, Inc. sent a letter dated July 21, 2011 entitled "URGENT MEDICAL DEVICE CORRECTION" to all affected customers. The letter included the name product, software versions, and reason for recall and informed them the problem is under investigation. In the meantime, customers are asked to wait 30 seconds after the finalization of the decontamination cycle before initiating the next operation, in order to avoid this issue. For questions on this recall please call Roche at (800) 428-5074.
- Initiated
- 2011-07-21
- Root cause
- Other
- Terminated
- 2012-07-02
- Distribution
- Nationwide Distribution; including the islands of Puerto Rico and Hawaii.
- Recall number
- Z-3078-2011
- Recall status
- Terminated
- CDRH record ID
- 102425
- FDA classified
- 2011-08-29
- Recall event number
- 59370