Skip to content

Public FDA record

Z-3081-2011: Mini MaxLock Extreme Plating System Case & Tray, Model Number(s): MXM-901, Product Usage: The Mini Max

Recall record Z-3081-2011 for Orthohelix Surgical Designs Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-3081-2011
Company in FDA record
Orthohelix Surgical Designs Inc
FDA classification date
Aug 29, 2011
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 17, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

The surgical tray does not meet our Instruction for Use (IFU) for dry time requirement.

Source facts

Action
Orthohelix Surgical Designs Inc sent a RECALL NOTIFICATION letter dated July 6, 2011 and an UPDATED URGENT RECALL 2ND NOTIFICATION letter dated July 13, 2011 to all affected customers. The letter identified the recalled product, the problem, and the Field Correction actions to be taken. Customers were instructed to remove the blue rubber mat in the utility bay of the tray base immediately, tag the mat with the tray number, and return it to recalling firm immediately. The letter states once customers have removed the blue rubber mat from the tray base, they may continue to use the tray and system normally. No further actions or precautions are needed. Distributors were also asked to notify their customers of the recall. If you have any questions, please call 330-247-1444 or 330-247-1445.
Initiated
2011-07-06
Root cause
Nonconforming Material/Component
Terminated
2012-09-10
Distribution
USA Nationwide Distribution including the states of: AR, AZ, CA, CO, FL, GA, IL, IN, KY, MA, MD, MI, MO, NC, NE, NJ, NM, NY, OH, OR, PA, SD, TN, TX, UT, VA, WA, WI & WY.
Recall number
Z-3081-2011
Recall status
Terminated
CDRH record ID
102838
FDA classified
2011-08-29
Recall event number
59462

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗