Skip to content

Public FDA record

Z-3082-2026: LUX-Dx, LUX-Dx II, LUX-Dx II+ Insertable Cardiac Monitor System, SERVER SW LAT CLARITY DS EU, Model Numbe

Recall record Z-3082-2026 for Boston Scientific Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-3082-2026
Company in FDA record
Boston Scientific Corporation
FDA classification date
Sep 2, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

There is potential for a delay in AF Burden Alert notification affecting the server software which supports the Insertable Cardiac Monitors (ICMs). Delayed notification of an AF Burden Alert may delay clinical evaluation or therapeutic intervention of an atrial arrhythmia.

Source facts

Action
An Urgent Medical Device Correction dated August 2026 was sent to customers. Boston Scientific is notifying you of the potential for a delay in AF Burden Alert notification affecting the LATITUDE" Clarity server software which supports the LUX-Dx" family of ICMs. This delayed AF Burden Alert notification behavior by LATITUDE Clarity may occur if the Alert for AF Burden e Threshold is configured to ">0 hours per day" setting (a non-default configuration). The ICM is performing as intended with continued AF detection, data storage, and transmission of device data. Boston Scientific has received one complaint related to this delayed notification behavior by the LATITUDE Clarity System. No patient injuries or deaths have been reported. A server software update is planned to correct this behavior. Configuration Options for Atrial Alert Notifications: Atrial arrhythmia alert notification preferences may vary, and AF configurations may be set for individual patients and/or group settings. Until updated server software is available, consider the following options: Change Threshold for AF Burden Alert Setting Enable AF Events Alert with Minimum Duration If Unchanged, Potential Delay in AF Burden Alert Notification " Boston Scientific does not recommend replacement of the implanted ICM. " Please share this notification with all HCPs within your organization who manage patients using the LUXDx ICM and LATITUDE Clarity System. Mitigation. The LATITUDE Clarity system is designed to detect atrial fibrillation through two independent mechanisms: AF Burden and AF Event Alerts. Until a software update to the LATITUDE Clarity server software is available, configuring AF Events or changing AF Burden Alerts settings can mitigate delays in full device interrogation and enable the LATITUDE Clarity system to display alert notification as intended. Additional Information. The Regulatory Authority of your country has been informed of this communication. Adverse events should be reported
Initiated
2026-08-06
Root cause
Software design
Distribution
Worldwide - US Nationwide distribution.
Recall number
Z-3082-2026
Recall status
Open, Classified
CDRH record ID
221753
FDA classified
2026-09-02
Recall event number
99524

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗