Skip to content

Public FDA record

Z-3083-2011: Fresenius 2008K@home The Fresenius 2008K@home is indicated for acute and chronic dialysis therapy in an

Recall record Z-3083-2011 for Fresenius Medical Care Holdings, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-3083-2011
Company in FDA record
Fresenius Medical Care Holdings, Inc.
FDA classification date
Aug 29, 2011
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 17, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Upgrade software to implement new on screen graphics and instructions, the addition of a bracket on the side of the machine to temporarily hold the bloodlines during set up in order to conform to approved 510(k)

Source facts

Action
Fresenius Medical NA issued a formal field correction notification dated May 4 ,2011, sent by certified mail with return receipt requested. The letter identified the product the problem and the action needed to be taken by the customer. The customer was notified of the retrofit that would be done to their machines. The patients' facilities were made aware of the changes that would be made to each of their patient's 2008K@home machines on May 4, 2011. On May 13, 2011, all facilities with active 2008K@home machines were sent a notification letter of the retrofit. Please return the receipt for confirmation that you have received this notice. Please call (781) 699-2053, to schedule a certified technician to conduct the machine retrofit.
Initiated
2011-05-04
Root cause
Device Design
Terminated
2014-08-14
Distribution
Nationwide Distribution. Canada - FMCNA is not conducting the recall to Canada, as Health Canada did not require the same modifications to the machines
Recall number
Z-3083-2011
Recall status
Terminated
CDRH record ID
102875
FDA classified
2011-08-29
Recall event number
59483

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.