Open original FDA record ↗Recall record
- FDA identifier
- Z-3084-2011
- Company in FDA record
- Maquet Inc.
- FDA classification date
- Aug 29, 2011
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Maquet HL 20 Twin Pump Modules may unintentionally increase speeds for a 1.5-2 second period of time before returning to its original speed.
Source facts
- Action
- Maquet sent a PRODUCT RECALL NOTIFICATION LETTER/return response forms, dated June 17, 2011, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to discontinue use of the affected product until a software update was performed by their trained technicians. Customers were also asked to forward the information contained in the notification letter to all users and to all appropriate Biomedical Staff. Customers were instructed to acknowledge receipt of the notification letter by completing and faxing back the attached response form to 973-807-9210 whether or not they have any of the affected product. For questions customers were instructed to contact their local MAQUET representative. For questions regarding this recall call 973-709-7994.
- Initiated
- 2011-06-20
- Root cause
- Device Design
- Terminated
- 2012-10-16
- Distribution
- Nationwide Distribution
- Recall number
- Z-3084-2011
- Recall status
- Terminated
- CDRH record ID
- 103140
- FDA classified
- 2011-08-29
- Recall event number
- 59577