Open original FDA record ↗Recall record
- FDA identifier
- Z-3084-2017
- Company in FDA record
- Ameditech Inc
- FDA record creation date
- Aug 29, 2017
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class III
Reason stated in the source
The kit box label incorrectly identifies PCP (instead of TCA) as one of the 10 drug analytes included in the kit. The pouch label, product insert, and the device itself are labeled correctly.
Source facts
- Action
- Ameditech, Inc. sent an Medical Device Labeling Correction letter to all affected consignees. Remaining product at the firm was quarantined and will be scrapped. Consignees will be notified of the labeling error with a correction notification. The kit label has been corrected and is available for future product. The firm plans to conduct effectiveness checks based upon receipt of the Label Correction notification by consignees. Firm indicates that no customer action is required, so no formal reply from distributors is necessary. Customers who have not had confirmed receipt of correction notification will be telephoned. For questions regarding this recall call 858-805-3607.
- Initiated
- 2017-06-26
- Root cause
- Labeling Change Control
- Terminated
- 2019-08-06
- Distribution
- Nationwide Distribution to AZ, IL, IN, KY, NV, PA, TN, and TX
- Recall number
- Z-3084-2017
- Recall status
- Terminated
- CDRH record ID
- 157665
- FDA record created
- 2017-08-29
- Recall event number
- 77702