Heightened probability of blood leak from the connection between the Arterial Alpha Clip and the main line or pump segment. Leak during treatment may lead to blood loss and hypovolemia.
Source facts
Action
On 8/10/2026, the firm began notifying consignees via URGENT MEDICAL DEVICE RECALL letters.
Customers were instructed to
1. Identify any affected product on hand.
2. IMMEDIATELY segregate and discontinue use of affected lot.
3. Contact the call center at 1-888-943-4901 and reference Event 8514, between
8:00 AM 4:30 PM Eastern daily, for return of affected lots and replacement of your
stock.
4, Complete and return the included reply form.
Initiated
2026-08-10
Root cause
Component change control
Distribution
Worldwide - US Nationwide distribution in the states of AL, CA, CO, FL, GA, IL, KY, LA, MA, MD, MN, MO, NC, NH, NH, NY, OH, PA, TN, TX, VA and the country of Lithuania.
Recall number
Z-3086-2026
Recall status
Open, Classified
CDRH record ID
221854
FDA classified
2026-09-02
Recall event number
99556